Monday, 7 March 2016

PWM Lasers



            Many lasers, especially CO2 lasers, use pulse width modulation (PWM) to control the power level of the laser. This is not true, pulsed operation, but rather a reduction of the duty cycle to lower the average power. The beam operates as if it were CW, and many operators do not even realize that the laser is pulsing. However, when attempting to measure a PWM laser with a scanning slit profiler, it must be treated as a pulsed laser source. To use the pulsed mode of the NanoScan the laser's pulse frequency must be at least several kHz, and the combination of the frequency and beam size must provide a sufficient number of pulses across the beam to generate a meaningful profile. Eight to ten pulses are a reasonable minimum. PWM lasers usually operate around 10kHz. The relationship of the beam size and frequency is a fairly simple mathematical model. The NanoScan drum speed is software controlled from 1.25Hz to 20Hz. There are two available drum sizes for the NanoScan; the standard head has a drum diameter of 42mm and the large aperture and high power heads use a larger drum with 84mm diameter. On the 42mmdrum at the 1.25Hz rotation rate the slits travel at around 116.6mm per second or 116.6μm per millisecond. At a 10kHz laser repetition rate, a 175μm beam would have 15 pulses during the time that the slit was traversing it. This would provide enough data to generate a meaningful profile. A smaller beam would require a faster pulse rate, a larger one could perhaps run at a lower repetition rate. For example, a 1.0mm beam could be measured with a pulse rate as low as 2kHz and still provide a profile.